Record details

Title
    A Simple Contactless Conductivity Detector Employing a Medium Wave Radio Integrated Circuit for the Signal Treatment.
Author
    Fenclová, Kateřina
    Opekar, František
    Štulík, Karel
Language
    anglicky
Publication type
    článek v periodiku
Source title - serial
    Electroanalysis
Vol./nr.
    Roč. 22, č. 2
Pages
    7
Year
    2010
Thesaurus term
    AM receiver integrated circuit
    Conductometry
    Contactless conductivity detection
    Detection in flowing liquids
    Drinking waters.
    TDA 1072A
Keyword
    Circuit
    Conductivity
    Contactless
    Detector
    Employing
    Integrated
    Medium
    Radio
    Signal
    Simple
    Treatment
    Wave
Abstract (in english)
   A new procedure has been tested for the treatment of the alternating signal coming from a contactless conductivity detection cell. An unmodulated or an amplitude-modulated a.c. voltage is applied to the cell and the a.c. current passing through the cell is treated by a TDA 1072A integrated circuit, obtaining a signal depending on the conductivity of the quiescent or flowing solution inside the cell. The detector was used to measure the conductivities of various drinking waters and the values obtained were in a good agreement with those provided by a commercial contact conductometer.